Configuration File¶
This is the core of the algorithm, so this file has to be filled properly based on your data. Even if all key parameters of the algorithm are listed in the file, only few are likely to be modified by a non-advanced user. The configuration file is divided in several sections. For all those sections, we will review the parameters, and tell you what are the most important ones
Data¶
The data section is:
file_format = # Can be raw_binary, openephys, hdf5, ... See >> spyking-circus help -i for more info
stream_mode = None # None by default. Can be multi-files, or anything depending to the file format
mapping = ~/probes/mea_252.prb # Mapping of the electrode (see http://spyking-circus.rtfd.ord)
suffix = # Suffix to add to generated files
global_tmp = True # should be False if local /tmp/ has enough space (better for clusters)
overwrite = True # If you want to filter or remove artefacts on site. Data are duplicated otherwise
output_dir = # By default, generated data are in the same folder as the data.
Warning
This is the most important section, that will allow the code to properly load your data. If not properly filled, then results will be wrong. Note that depending on your file_format, you may need to add here several parameters, such as sampling_rate
, data_dtype
, … They will be requested if they can not be infered from the header of your data structure. To check if data are properly loaded, consider using the preview mode before launching the whole algorithm
- Parameters that are most likely to be changed:
file_format
You must select a supported file format (see What are the supported formats) or write your own wrapper (see Write your own data format)mapping
This is the path to your probe mapping (see How to design a probe file)global_temp
If you are using a cluster with NFS, this should be False (local /tmp/ will be used by every nodes)stream_mode
If streams in you data (could be multi-files, or even in the same file) should be processed together (see Using multi files)overwrite
If True, data are overwritten during filtering, assuming the file format has write access. Otherwise, an external raw_binary file will be created during the filtering step, if any.ouput_dir
If you want all the file generated by SpyKING CIRCUS to be in a particular directory, instead of next to the raw data
Detection¶
The detection section is:
radius = auto # Radius [in um] (if auto, read from the prb file)
N_t = 5 # Width of the templates [in ms]
spike_thresh = 6 # Threshold for spike detection
peaks = negative # Can be negative (default), positive or both
matched-filter = False # If True, we perform spike detection with matched filters
matched_thresh = 5 # Threshold for detection if matched filter is True
alignment = True # Realign the waveforms by oversampling
- Parameters that are most likely to be changed:
N_t
The temporal width of the templates. For in vitro data, 5ms seems a good value. For in vivo data, you should rather use 3 or even 2msradius
The spatial width of the templates. By default, this value is read from the probe file. However, if you want to specify a larger or a smaller value [in um], you can do it herespike_thresh
The threshold for spike detection. 6-7 are good valuespeaks
By default, the code detects only negative peaks, but you can search for positive peaks, or bothmatched-filter
If activated, the code will detect smaller spikes by using matched filteringmatched_thresh
During matched filtering, the detection thresholdalignment
By default, during clustering, the waveforms are realigned by oversampling at 5 times the sampling rate and using bicubic spline interpolation
Filtering¶
The filtering section is:
cut_off = 300, auto # Min and Max (auto=nyquist) cut off frequencies for the band pass butterworth filter [Hz]
filter = True # If True, then a low-pass filtering is performed
remove_median = False # If True, median over all channels is substracted to each channels (movement artefacts)
Warning
The code performs the filtering of your data writing on the file itself. Therefore, you must
have a copy of your raw data elsewhere. Note that as long as your keeping the parameter files, you can relaunch the code safely: the program will not filter multiple times the data, because of the flag filter_done
at the end of the configuration file.
- Parameters that are most likely to be changed:
cut_off
The default value of 500Hz has been used in various recordings, but you can change it if needed. You can also specify the upper bound of the Butterworth filterfilter
If your data are already filtered by a third program, turn that flag to Falseremove_median
If you have some movement artefacts in your in vivo recording, and want to substract the median activity over all analysed channels from each channel individually
Triggers¶
The triggers section is:
trig_file = # External stimuli to be considered as putative artefacts [in trig units] (see documentation)
trig_windows = # The time windows of those external stimuli [in trig units]
trig_unit = ms # The unit in which times are expressed: can be ms or timestep
clean_artefact = False # If True, external artefacts induced by triggers will be suppressed from data
dead_file = # Portion of the signals that should be excluded from the analysis [in dead units]
dead_unit = ms # The unit in which times for dead regions are expressed: can be ms or timestep
ignore_times = False # If True, any spike in the dead regions will be ignored by the analysis
make_plots = # Generate sanity plots of the averaged artefacts [Nothing or None if no plots]
- Parameters that are most likely to be changed:
trig_file
The path to the file where your artefact times and labels. See how to deal with stimulation artefactstrig_windows
The path to file where your artefact temporal windows. See how to deal with stimulation artefactsclean_artefact
If you want to remove any stimulation artefacts, defined in the previous files. See how to deal with stimulation artefactsmake_plots
The default format to save the plots of the artefacts, one per artefact, showing all channels. You can set it to None if you do not want anytrig_unit
If you want times/duration in thetrig_file
andtrig_windows
to be in timestep or msdead_file
The path to the file where the dead portions of the recording, that should be excluded from the analysis, are specified. . See how to deal with stimulation artefactsdead_unit
If you want times/duration in thedead_file
to be in timestep or msignore_times
If you want to remove any dead portions of the recording, defined indead_file
. See how to deal with stimulation artefacts
Whitening¶
The whitening section is:
chunk_size = 60 # Size of the data chunks [in s]
safety_time = 1 # Temporal zone around which templates are isolated [in ms]
temporal = False # Perform temporal whitening
spatial = True # Perform spatial whitening
max_elts = 10000 # Max number of events per electrode (should be compatible with nb_elts)
nb_elts = 0.8 # Fraction of max_elts that should be obtained per electrode [0-1]
output_dim = 5 # Can be in percent of variance explain, or num of dimensions for PCA on waveforms
- Parameters that are most likely to be changed:
output_dim
If you want to save some memory usage, you can reduce the number of features kept to describe a waveform.chunk_size
If you have a very large number of electrode, and not enough memory, you can reduce it
Clustering¶
The clustering section is:
extraction = median-raw # Can be either median-raw (default), median-pca, mean-pca, mean-raw, or quadratic
safety_space = True # If True, we exclude spikes in the vicinity of a selected spikes
safety_time = 1 # Temporal zone around which templates are isolated [in ms]
max_elts = 10000 # Max number of events per electrode (should be compatible with nb_elts)
nb_elts = 0.8 # Fraction of max_elts that should be obtained per electrode [0-1]
nclus_min = 0.002 # Min number of elements in a cluster (given in percentage)
max_clusters = 10 # Maximal number of clusters for every electrodes
nb_repeats = 3 # Number of passes used for the clustering
make_plots = # Generate sanity plots of the clustering
sim_same_elec = 3 # Distance within clusters under which they are re-merged
cc_merge = 0.975 # If CC between two templates is higher, they are merged
dispersion = (5, 5) # Min and Max dispersion allowed for amplitudes [in MAD]
smart_search = True # Parameter to activate the smart search mode
smart_select = False # Experimental: activate the smart selection of centroids (max_clusters is ignored)
noise_thr = 0.8 # Minimal amplitudes are such than amp*min(templates) < noise_thr*threshold
remove_mixture = True # At the end of the clustering, we remove mixtures of templates
Note
This is the a key section, as bad clustering will implies bad results. However, the code is very robust to parameters changes.
- Parameters that are most likely to be changed:
extraction
The method to estimate the templates.Raw
methods are slower, but more accurate, as data are read from the files.PCA
methods are faster, but less accurate, and may lead to some distorted templates.Quadratic
is slower, and should not be used.max_elts
The number of elements that every electrode will try to collect, in order to perform the clusteringnclus_min
If you have too many clusters with few elements, you can increase this value. This is expressed in percentage of collected spike per electrode. So one electrode collecting max_elts spikes will keep clusters with more than nclus_min.max_elts. Otherwise, they are discardedmax_clusters
This is the maximal number of cluster that you expect to see on a given electrode. For in vitro data, 10 seems to be a reasonable value. For in vivo data and dense probes, you should set it to 10-15. Increase it only if the code tells you so.nb_repeats
The number of passes performed by the algorithm to refine the density landscapesmart_search
By default, the code will collect only a subset of spikes, randomly, on all electrodes. However, for long recordings, or if you have low thresholds, you may want to select them in a smarter manner, in order to avoid missing the large ones, under represented. If the smart search is activated, the code will first sample the distribution of amplitudes, on all channels, and then implement a rejection algorithm such that it will try to select spikes in order to make the distribution of amplitudes more uniform. This can be very efficient, and may become True by default in future releases.smart_select
This option (experimental) should boost the quality of the clustering, by selecting the centroids in a automatic manner. If activated themax_clusters
parameter is ignoredcc_merge
After local merging per electrode, this step will make sure that you do not have duplicates in your templates, that may have been spread on several electrodes. All templates with a correlation coefficient higher than that parameter are merged. Remember that the more you merge, the faster is the fitdispersion
The spread of the amplitudes allowed, for every templates, around the centroid.remove_mixture
By default, any template that can be explained as sum of two others is deleted.make_plots
By default, the code generates sanity plots of the clustering, one per electrode.
Fitting¶
The fitting section is:
chunk = 1 # Size of chunks used during fitting [in second]
gpu_only = True # Use GPU for computation of b's AND fitting
amp_limits = (0.3, 30) # Amplitudes for the templates during spike detection
amp_auto = True # True if amplitudes are adjusted automatically for every templates
max_chunk = inf # Fit only up to max_chunk
collect_all = False # If True, one garbage template per electrode is created, to store unfitted spikes
- Parameters that are most likely to be changed:
chunk
again, to reduce memory usage, you can reduce the size of the temporal chunks during fitting. Note that it has to be one order of magnitude higher than the template widthN_t
gpu_only
By default, all operations will take place on the GPU. However, if not enough memory is available on the GPU, then you can turn this flag to False.max_chunk
If you just want to fit the first N chunks, otherwise, the whole file is processedcollect_all
If you want to also collect all the spike times at which no templates were fitted. This is particularly useful to debug the algorithm, and understand if something is wrong on a given channel
Merging¶
The merging section is:
cc_overlap = 0.75 # Only templates with CC higher than cc_overlap may be merged
cc_bin = 1 # Bin size for computing CC [in ms]
correct_lag = False # If spikes are aligned when merging. May be better for phy usage
auto_mode = 0 # If >0, merging will be automatic (see doc, 1e-05 is a good value) [0-1]
- To know more about how those merges are performed and how to use this option, see Automatic Merging. Parameters that are most likely to be changed:
correct_lag
By default, in the meta-merging GUI, when two templates are merged, the spike times of the one removed are simply added to the one kept, without modification. However, it is more accurate to shift those spike, in times, by the temporal shift that may exist between those two templates. This will lead to a better visualization in phy, with more aligned spikesauto_mode
If your recording is stationary, you can try to perform a fully automated merging. By setting a value between 0 and 1, you control the level of merging performed by the software. Values such as 1e-05 should be a good start, but see see Automatic Merging for more details.
Converting¶
The converting section is:
erase_all = True # If False, a prompt will ask you to export if export has already been done
sparse_export = False # If True, data for phy are exported in a sparse format. Need recent version of phy
export_pcs = prompt # Can be prompt [default] or in none, all, some
export_all = False # If True, unfitted spikes will be exported as the last Ne templates
- Parameters that are most likely to be changed:
erase_all
If you want to always erase former export, and skip the promptsparse_export
If you have a large number of templates or a very high density probe, you should use the sparse format for phyexport_pcs
If you already know that you want to have all, some, or no PC and skip the promptexport_all
If you used thecollect_all
mode in the[fitting]
section, you can export unfitted spike times to phy. In this case, the last N templates, if N is the number of electrodes, are the garbage collectors.
Extracting¶
The extracting section is:
safety_time = 1 # Temporal zone around which spikes are isolated [in ms]
max_elts = 10000 # Max number of events per templates (should be compatible with nb_elts)
nb_elts = 0.8 # Fraction of max_elts that should be obtained per electrode [0-1]
output_dim = 5 # Percentage of variance explained while performing PCA
cc_merge = 0.975 # If CC between two templates is higher, they are merged
noise_thr = 0.8 # Minimal amplitudes are such than amp*min(templates) < noise_thr*threshold
This is an experimental section, not used by default in the algorithm, so nothing to be changed here
Validating¶
The validating section is:
nearest_elec = auto # Validation channel (e.g. electrode closest to the ground truth cell)
max_iter = 200 # Maximum number of iterations of the stochastic gradient descent (SGD)
learning_rate = 1.0e-3 # Initial learning rate which controls the step-size of the SGD
roc_sampling = 10 # Number of points to estimate the ROC curve of the BEER estimate
test_size = 0.3 # Portion of the dataset to include in the test split
radius_factor = 0.5 # Radius factor to modulate physical radius during validation
juxta_dtype = uint16 # Type of the juxtacellular data
juxta_thresh = 6 # Threshold for juxtacellular detection
juxta_valley = False # True if juxta-cellular spikes are negative peaks
juxta_spikes = # If none, spikes are automatically detected based on juxta_thresh
filter = True # If the juxta channel need to be filtered or not
make_plots = png # Generate sanity plots of the validation [Nothing or None if no plots]
Please get in touch with us if you want to use this section, only for validation purposes. This is an implementation of the BEER metric